How to Make EDX Analysis Achievable on a K-kit |
By pointing the window long side to the detector and tilting the holder at some angles, which could make the EDX analysis achievable on a k-kit. |
Without tilting | Give a tilt toward the EDX detector |
For some types of TEM installed with multiple EDX detectors, the X-ray signals excited would not entirely be blocked by the bulk body of k-kit. We usually can get it a clear signal from K-kit, no need to turn any of body rotation or tilting. |
| ||
The EDX signal of gold could be detected directly, no need to make any tilt at the K-kit body. The gold nanoparticles were formed from the reduction process of AuCl4 solution in K-kit, and we used FEI Talos TEM, which with four EDX dectors around the specimen area of the TEM column, in this test. |
EDX detectable areas of k-kit in different kinds of TEM equipment |
|
|
Example I TEM: Hitachi 7700EDX: Single Detector Liquid Sample: AuCl3 Tilt Angle: 0°, 10°, 20° Measured Area: Center |
|
Example II TEM: Hitachi 7700EDX: Single Detector Liquid Sample: AgNA Sample Prep.: Wet Mode Tilt Angle: 20° Measured Area: Left Site | (The location to be measured) |